print Print
Please select which sections you would like to print:
verifiedCite
While every effort has been made to follow citation style rules, there may be some discrepancies. Please refer to the appropriate style manual or other sources if you have any questions.
Select Citation Style
Feedback
Corrections? Updates? Omissions? Let us know if you have suggestions to improve this article (requires login).
Thank you for your feedback

Our editors will review what you’ve submitted and determine whether to revise the article.

The second major category of instrumental analysis is electroanalysis. The electroanalytical methods use electrically conductive probes, called electrodes, to make electrical contact with the analyte solution. The electrodes are used in conjunction with electric or electronic devices to which they are attached to measure an electrical parameter of the solution. The measured parameter is related to the identity of the analyte or to the quantity of the analyte in the solution.

The electroanalytical methods are divided into categories according to the electric parameters that are measured. The major electroanalytical methods include potentiometry, amperometry, conductometry, electrogravimetry, voltammetry (and polarography), and coulometry. The names of the methods reflect the measured electric property or its units. Potentiometry measures electric potential (or voltage) while maintaining a constant (normally nearly zero) electric current between the electrodes. Amperometry monitors electric current (amperes) while keeping the potential constant. Conductometry measures conductance (the ability of a solution to carry an electric current) while a constant alternating-current (AC) potential is maintained between the electrodes. Electrogravimetry is a gravimetric technique similar to the classical gravimetric methods that were described above, in which the solid that is weighed is deposited on one of the electrodes. Voltammetry is a technique in which the potential is varied in a regular manner while the current is monitored. Polarography is a subtype of voltammetry that utilizes a liquid metal electrode. Coulometry is a method that monitors the quantity of electricity (coulombs) that are consumed during an electrochemical reaction involving the analyte.

Most of the electroanalytical methods rely on the flow of electrons between one or more of the electrodes and the analyte. The analyte must be capable of either accepting one or more electrons (known as reduction) from the electrode or donating one or more electrons (oxidation) to the electrode. As an example, ferric iron (Fe3+) can be assayed because it can undergo a reduction to ferrous iron (Fe2+) by accepting an electron from the electrode as shown in the following reaction:

Equation.

Conductometry

This is the method in which the capability of the analyte to conduct an electrical current is monitored. From Ohm’s law (E = IR) it is apparent that the electric current (I) is inversely proportional to the resistance (R), where E represents potential difference. The inverse of the resistance is the conductance (G = 1/R). As the conductance of a solution increases, its ability to conduct an electric current increases.

In liquid solutions current is conducted between the electrodes by dissolved ions. The conductance of a solution depends on the number and types of ions in the solution. Generally small ions and highly charged ions conduct current better than large ions and ions with a small charge. The size of the ions is important because it determines the speed with which the ions can travel through the solution. Small ions can move more rapidly than larger ones. The charge is significant because it determines the amount of electrostatic attraction between the electrode and the ions.

Because conductometric measurements require the presence of ions, conductometry is not useful for the analysis of undissociated molecules. The measured conductance is the total conductance of all the ions in the solution. Since all ions contribute to the conductivity of a solution, the method is not particularly useful for qualitative analysis—i.e., the method is not selective. The two major uses of conductometry are to monitor the total conductance of a solution and to determine the end points of titrations that involve ions. Conductivity meters are used in conjunction with water purification systems, such as stills or deionizers, to indicate the presence or absence of ion-free water.

Conductometric titration curves are prepared by plotting the conductance as a function of the volume of added titrant. The curves consist of linear regions prior to and after the end point. The two linear portions are extrapolated to their point of intersection at the end point. As in other titrations, the end-point volume is used to calculate the amount or concentration of analyte that was originally present.

Voltammetry

Voltammetry can be used for both qualitative and quantitative analysis of a wide variety of molecular and ionic materials. In this method, a set of two or three electrodes is dipped into the analyte solution, and a regularly varying potential is applied to the indicator electrode relative to the reference electrode. The analyte electrochemically reacts at the indicator electrode. The reference electrode is constructed so that its potential is constant regardless of the solution into which it is dipped. Usually a third electrode (an auxiliary or counter electrode) is placed in the solution for the purpose of carrying most of the current. The potential is controlled between the indicator electrode and the reference electrode, but the current flows between the auxiliary electrode and the indicator electrode.

Classic polarography

The several forms of voltammetry differ in the type of varying potential that is applied to the indicator electrode. Polarography is voltammetry in which the indicator electrode is made of mercury or, rarely, another liquid metal. In classic polarography, mercury drops from a capillary tube. The surface of the mercury drop is the site of the electrochemical reaction with the analyte. The manner in which the direct-current (DC) potential of the indicator electrode varies with time is a potential (or voltage) ramp. In the most common case, the potential varies linearly with time, and the analytical method is known as linear sweep voltammetry (LSV).

Typically the potential is initially adjusted to a value at which no electrochemical reaction occurs at the indicator electrode. The potential is scanned in a direction that makes an electrochemical reaction more favourable. If reduction reactions are studied, the electrode is made more cathodic (negative); if oxidations are studied, the electrode is made more anodic (positive). Initially the current that is measured, before the electrochemical reaction begins, is small. As the electrode potential is changed, however, sufficient energy is applied to the indicator electrode to cause the reaction to take place. As the reaction occurs, electrons are withdrawn from the electrode (for electrochemical reductions) or donated to the electrode (for oxidations), and a current flows in the external electrical circuit. A voltammogram is a plot of the current as a function of the applied potential. The shape of a voltammogram depends on the type of indicator electrode and the potential ramp that are used. In nearly all cases, the voltammogram has a current wave as shown in Figure 1 or a current peak as shown in Figure 2.

This technique can be used for qualitative analysis because substances exhibit characteristic peaks or waves at different potentials. The height (current) of the wave or the peak, as measured by extrapolating the linear portion of the curve prior to the wave or peak and taking the difference between this extrapolated line and the current peak or plateau, is directly proportional to the concentration of the analyte and can be used for quantitative analysis. Normally the concentration corresponding to the peak or wave height of the analyte is determined from a working curve.

Triangular wave voltammetry

Triangular wave voltammetry (TWV) is a method in which the potential is linearly scanned to a value past the potential at which an electrochemical reaction occurs and is then immediately scanned back to its original potential. A triangular wave voltammogram usually has a current peak on the forward scan and a second, inverted peak on the reverse scan representing the opposite reaction (oxidation or reduction) to that observed on the forward scan. Cyclic voltammetry is identical to TWV except in having more than one cycle of forward and reverse scans successively completed.

AC voltametry

During AC voltammetry an alternating potential is added to the DC potential ramp used for LSV. Only the AC portion of the total current is measured and plotted as a function of the DC potential portion of the potential ramp. Because flow of an alternating current requires the electrochemical reaction to occur in the forward and reverse directions, AC voltammetry is particularly useful for studying the extent to which electrochemical reactions are reversible.

Pulse and differential pulse voltammetry

Differential pulse voltammetry adds a periodically applied potential pulse (temporary increase in potential) to the voltage ramp used for LSV. The current is measured just prior to application of the pulse and at the end of the applied pulse. The difference between the two currents is plotted as a function of the LSV ramp potential. Pulse voltammetry utilizes a regularly increasing pulse height that is applied at periodic intervals. In pulse and differential pulse polarography the pulses are applied just before the mercury drop falls from the electrode. Typically the pulse is applied for about 50–60 milliseconds; and the current is measured during the last 17 milliseconds of each pulse. The voltammogram is a plot of the measured current as a function of the potential of the pulse. Many other variations of voltammetry also are available but are not as commonly used. Sketches showing the various potential ramps that are applied to the indicator electrode during the various types of polarography, along with the typical corresponding polarograms, are shown in Figure 3.